Evaluation of sound pressure levels in a pediatric intensive care unit

Open Science Journal

View Publication Info
 
 
Field Value
 
Title Evaluation of sound pressure levels in a pediatric intensive care unit
 
Creator Mendes, Caroline de Araújo
da Silva, Carlos Roberto Lyra
de Souza, Luciane Velasque
Bridi, Adriana Carla
Cabral, Joice Alves
Dutra, Luana Borges
Barreiros, Raphael Neves
Garcia, Andrea dos Santos
Salú, Margarida dos Santos
Barroso, Natália
 
Subject Nursing
Noise Meters; Noise; Intensive care; Pediatrics
 
Description This study objectives were to measure the sound pressure levels found in the pediatric intensive care unit in a federal institution of Rio de Janeiro; to verify differences in noise levels during the morning and afternoon; to confront the sound pressure levels found against acceptable levels according to national and international noise organizations; to count the quantity of alarms triggered by each type of medical care equipment selected (multiparameter monitor, mechanical ventilator and infusion pump); to verify the relevance in the scientific world about pediatric intensive care unit noise through bibliometrics and to address the trinomial care technology - noise - implications on care. It’s an observational, exploratory, quantitative study, organized in three steps: Parameter collection and decibel meter calibration - data were based on the study by Salú, et al (2015) ; Data collection: 40 hours of discontinued observation (8am to 16pm) on different days for a period two months using two decibel meters; Data processing: An Excel spreadsheet was created for the database and data analysis was performed with the help of Microsoft Office Excel 2010 and Program R, organized into graphs and tables. 61% of the alarms corresponded to the mechanical ventilator; Bed E had the lowest standard deviation (SD = 2.945) and the highest median (69.5dBA). Even by removing the E bed from the analysis, there is a significant difference (p <0.001) between sound pressure levels. The median of the afternoon (28.2dBA); and morning (26.1dBA). Mechanical fan and monitor generated higher sounds; the pediatric intensive care unit has considerably exceeded that recommended by national and international noise organizations; afternoon generated higher sounds than morning. Keywords: Noise Meters; Noise; Intensive care; Pediatrics
 
Publisher Open Science Journal
 
Contributor
 
Date 2020-05-15
 
Type info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion

 
Format application/pdf
 
Identifier https://osjournal.org/ojs/index.php/OSJ/article/view/2415
10.23954/osj.v5i2.2415
 
Source Open Science Journal; Vol 5, No 2 (2020): Open Science Journal
2466-4308
 
Language eng
 
Relation https://osjournal.org/ojs/index.php/OSJ/article/view/2415/275
https://osjournal.org/ojs/index.php/OSJ/article/downloadSuppFile/2415/1137
 
Rights Copyright (c) 2020 Open Science Journal
http://creativecommons.org/licenses/by/4.0
 

Contact Us

The PKP Index is an initiative of the Public Knowledge Project.

For PKP Publishing Services please use the PKP|PS contact form.

For support with PKP software we encourage users to consult our wiki for documentation and search our support forums.

For any other correspondence feel free to contact us using the PKP contact form.

Find Us

Twitter

Copyright © 2015-2018 Simon Fraser University Library